Pharmaceutical Chemistry — 05
Analytical procedure
Quantifying acetylcysteine in the 20 % w/v nebulizing solution by reverse-phase HPLC, and assessing oxidative related substances against British Pharmacopoeia limits.
Objective. To accurately quantify the acetylcysteine content of the 20 % w/v solution using High Performance Liquid Chromatography (HPLC), and assess the presence of any related substances (oxidative degradation products) against the acceptance limits found in the British Pharmacopoeia.
Formulation
20 % w/v
Batch size
100 ml
Primary assay
RP-HPLC, 205 nm
Cross-check
Iodometric
5.1
Introduction to the analytical procedure
An analytical procedure can be described as a step-by-step process that is used to qualify (identify), quantify or characterise a certain chemical substance or mixture (American Chemical Society, n.d.). In this case, we are qualifying and quantifying acetylcysteine in the 20 % w/v nebulizing solution.
To qualify and quantify acetylcysteine, High-Performance Liquid Chromatography (HPLC) is used as the primary assay. This is an analytical technique used to separate, identify and therefore quantify specific individual components in a liquid mixture. The process is carried out by passing a mobile phase through a column that is tightly packed under high pressure. The acetylcysteine nebulizing solution is treated as a liquid mixture, and the aim is to separate, identify and quantify the acetylcysteine content present.
The principle
A method of chromatographic separation based on the different ways in which species distribute between two non-miscible phases — the stationary phase and the mobile phase (British Pharmacopoeia Commission, 2022).
This method uses reversed-phase chromatography, meaning the stationary phase is non-polar and the mobile phase is polar. The stationary phase consists of octadecylsilyl silica gel (a C18-bonded silica), a non-polar, chemically modified material tightly packed within the column. The mobile phase is a polar aqueous mixture of methanol and buffered ammonium sulphate solution, flowing consistently through the system.
Separation occurs based on polarity: more polar species interact less strongly with the non-polar stationary phase and are carried through the column more quickly by the polar mobile phase, while less polar species are retained longer. As the sample passes through the column, molecules with greater affinity for the stationary phase move slowly, while those with greater affinity for the mobile phase move more quickly. As a result of these different speeds, the different species separate from one another and reach the detector at different times.
5.2
Materials required
Reference standards & reagents
- Acetylcysteine standard solution
- Methanol (HPLC grade)
- Ammonium sulphate
- Sodium pentanesulphonate
- Hydrochloric acid (2 M, for mobile phase pH adjustment)
- Purified water
For the iodometric cross-validation
- Glacial acetic acid
- Iodine volumetric solution (0.05 M iodine VS)
Consumables
- HPLC grade syringe filters (0.45 μm)
- Amber HPLC glass vials
- Volumetric flasks (Grade A): 10 ml, 100 ml
- Class A bulb pipettes (1 ml, 5 ml as required for standard / sample dilution)
- Clean, dry syringes for sample transfer and filtration
Calibration standards
A bracketing series of calibration standards, prepared from the acetylcysteine CRM stock solution and diluted with the mobile phase, spanning 80–120 % of the target working concentration (0.2 % w/v acetylcysteine, or 2000 μg/ml).
| % of target | Concentration (μg/ml) |
|---|---|
| 80 % | 1600 |
| 90 % | 1800 |
| 100 % | 2000 |
| 110 % | 2200 |
| 120 % | 2400 |
HPLC equipment
A pumping system
Delivers the mobile phase at a known and controlled flow rate. In HPLC, the mobile phase is pumped under high pressures.
An injector
The sample solution enters the mobile phase at the column head through the injection. This injection can be used under high pressures, making it suited to HPLC.
A chromatographic column
Holds the mobile and stationary phases, and is where separation of the species occurs. These columns must be controlled by temperature regulators, since temperature affects retention time, mobile phase viscosity, column pressure and the final sharpness of the peak.
A detector
More than one type may be used; the UV/Vis spectrophotometer is most common, though fluorescence spectrophotometers, electrochemical detectors and others also exist. The detector registers molecules leaving the column at different times and generates an electrical signal.
Data system
Turns the electrical signals from the detector into a graph with peaks representing the different species in the sample. This graph is known as a chromatogram.
Key HPLC terms
Retention time
The exact time taken for a species to reach the detector once injected. Most species have the same retention time under the same specific conditions — acetylcysteine will have the same retention time whenever HPLC is run under identical conditions, and it is through this that qualification of acetylcysteine is done.
Area of the peak
The peak area corresponds to the concentration of a species in the sample, enabling quantification.
Isocratic vs. gradient
Isocratic elution keeps the mobile phase constant for the duration of the run. Gradient elution changes the solvent mixture during the run in order to speed up the slower species.
5.3
HPLC conditions
| Stationary phase | A stainless steel column (25 cm × 5 mm), packed with octadecylsilyl silica gel for chromatography (5 μm particle size) |
| Elution | Isocratic |
| Mobile phase | 10 volumes of methanol and 90 volumes of 0.5 % w/v ammonium sulphate containing 0.02 M sodium pentanesulphonate; adjusted to pH 2.0 using 2 M hydrochloric acid |
| Flow rate | 1 ml per minute |
| Detection wavelength | 205 nm (UV/Vis) |
| Injection volume | 20 μL per solution |
| Run time | Three times the retention time of acetylcysteine |
5.4
Procedure for the standard solution
It is important to have a standard against which a comparison can be made, to ensure reliability and consistency — it provides a reference point when performing any procedure. Creation of a master stock solution is therefore required.
Procedure
- 01Accurately weigh 40.0 mg of acetylcysteine CRM.
- 02Transfer to a 10.0 ml volumetric flask and dissolve in a small amount of mobile phase.
- 03Dilute to volume with mobile phase to give a 4.0 mg/ml (0.4 % w/v, 4000 μg/ml) stock solution.
- 04Prepare fresh. Do not store beyond the working session, given the oxidation risk discussed in Section 4.3.
After the creation of this stock solution, the calibration curve is built using the equation C₁V₁ = C₂V₂. For the volume of stock solution required and the working detail for each target concentration, see the table under Section 4.7.
5.5
Procedure for sample preparation
The acetylcysteine nebulizing solution contains 200 mg/ml of acetylcysteine (20 % w/v). According to the BP monograph for Acetylcysteine Injection, the target concentration is 0.2 % w/v (2000 μg/ml). Using C₁V₁ = C₂V₂, a volume of 1.00 ml of the nebulizing solution is required, as shown in the calculations under Section 4.5.
Procedure
- 01Accurately pipette 1.00 ml of the 20 % w/v nebulizing solution into a 100 ml Class A volumetric flask using a Class A bulb pipette.
- 02Dilute to volume with mobile phase and invert at least 10 times to ensure homogeneity (do not shake vigorously, to avoid introducing oxygen or air into the solution).
- 03Resulting in a concentration of 2 mg/ml, which is equal to 2000 μg/ml and 0.2 % w/v.
Because this working concentration matches what is prescribed in the BP, the sample sits at the same concentration as the calibration midpoint. This minimises extrapolation error when reading off the final calibration curve.
5.6
System suitability and limits
System suitability is important in HPLC as it ensures the entire setup of the procedure is working properly and meets the criteria of the BP before testing the samples prepared above. It identifies issues that could exist with the equipment or the process before actual testing occurs, and in the end saves time and resources.
According to the British Pharmacopoeia for Acetylcysteine Injection, when detecting impurities caused by the oxidative degradation products of acetylcysteine, the HPLC test being performed is not valid unless it meets the following criteria:
- —In the chromatogram obtained with the system suitability solution, the height of the trough separating the cysteine and cystine peaks is less than one quarter of the height of the cysteine peak.
- —A peak corresponding to N,N′-diacetylcystine (retention time approximately 13 minutes) appears, with an area greater than the area of any corresponding peak in the chromatogram obtained from the freshly prepared system suitability solution — demonstrating the system's ability to detect this degradation product as it forms.
As with any procedure, limits exist. Limits explain the reliability and accuracy of this particular method; being aware of them ensures there is an understanding of whether the method is actually able to detect any impurities present, qualify and quantify the active pharmaceutical ingredient, and meet the required standards. Limitations related to the detection of impurities, according to the British Pharmacopoeia for Acetylcysteine Injection, are as follows:
≤ 1 %
The area of the N,N′-diacetylcystine peak is not greater than 1 % of the area of the acetylcysteine peak.
≤ 0.5 %
The area of the cysteine or cystine peak is not greater than 0.5 % of the corresponding peak area.
Σ ≤ 1 %
The sum of the areas of any secondary peaks, expressed against the area of the acetylcysteine peak.
< 0.1 %
Any peaks with an area less than 0.1 % of the acetylcysteine peak are disregarded.
5.7
Calibration curve construction
Once the calibration standards have been prepared, the calibration curve is constructed as follows:
- 01Inject 20 μL of each calibration standard into the HPLC system, using the chromatographic conditions specified in Section 5.3.
- 02Record the peak area corresponding to each calibration standard.
- 03Plot the recorded peak areas against their known concentrations to construct the calibration curve, and determine the regression line.
The calibration curve serves as the reference against which the sample's peak area is later compared to determine acetylcysteine content.
5.8
Sample analysis
Once the sample dilution is prepared, it is analysed as follows:
- 01Inject 20 μL of the prepared sample dilution into the HPLC system, under the same chromatographic conditions used for the calibration standards.
- 02Record the peak area corresponding to acetylcysteine, together with any other secondary peaks corresponding to cysteine, cystine or N,N′-diacetylcystine.
- 03Confirm the identity of the acetylcysteine peak by comparing its retention time to that of the calibration standards.
5.9
Data analysis
Quantification of acetylcysteine content is calculated by comparing the sample's peak area against the calibration curve obtained in Section 5.7.
- —Read the sample's acetylcysteine concentration directly from the calibration curve.
- —Calculate the % w/v content, accounting for the dilution factor applied during sample preparation, and compare this to the declared 20 % w/v label claim.
- —Assess related substances by expressing the area of each secondary peak (cysteine, cystine or N,N′-diacetylcystine) as a percentage of the acetylcysteine peak, and compare each against the corresponding limit specified in Section 5.6.
- —Confirm that all system suitability and limit criteria have been satisfied before accepting the result as valid.
5.10
Reporting
Acetylcysteine content is reported as % w/v of the nebulizing solution, alongside the % of label claim. Any related substances detected are reported as a percentage of the acetylcysteine peak area, referenced against the British Pharmacopoeia limits for Acetylcysteine Injection (British Pharmacopoeia Commission, 2022).
5.11
Cross-validation using iodometric titration
Acetylcysteine is a derivative of the amino acid L-cysteine and possesses a sulfhydryl group (-SH). Due to this additional hydrogen atom, acetylcysteine is in a reduced state and has additional electrons that can be donated. Iodine reacts readily with free thiol groups, oxidising them in a redox reaction. However, this reaction is not selective for acetylcysteine specifically: iodine will oxidise any free thiol present in the sample, including thiol-containing degradation products.
This lack of specificity is precisely why iodometric titration is unsuitable as the primary quantification method, and is instead used here as an independent cross-validation check against the HPLC result, following the British Pharmacopoeia's official assay method for Acetylcysteine Injection.
Procedure
- 01Measure out a volume of the nebulizing solution theoretically containing 0.4 g of acetylcysteine.
- 02Add 20 mL of glacial acetic acid to create the acidic environment required for the redox reaction.
- 03Titrate with 0.05 M iodine volumetric solution (VS).
- 04Iodine is continuously consumed by the free thiol groups present until no further reducible thiol remains. The next drop of iodine added is then in excess, producing a permanent pale yellow colour that marks the endpoint.
Titre equivalence
1 mL of 0.05 M iodine VS ≡ 16.23 mg acetylcysteine
British Pharmacopoeia Commission, 2022